What is the reason why the pressure gauge pionter does not move after pressurizing?

Mar 13, 2020

What is the reason why the pressure gauge pointer does not move after pressurizing?

 

This situation sometimes occurs when the pressure gauge is calibrated or used. When the pressure gauge is pressurized, the standard pressure gauge has moved or the piping system has pressure, but the working pressure gauge pointer does not move. What is the reason?

 

 

Analysis of the reasons why the pressure gauge pointer does not move:

 

1. It is possible that the valve under the pressure gauge has not been opened and the pressure gauge pointer does not move during pressurization;

2, The pressure guide hole under the pressure gauge connector is blocked, which is one of the reasons that the pointer of the pressure gauge does not move when pressurizing. In this case, as long as the position of the working pressure gauge is swapped with the position of the standard pressure gauge, it can be judged.

3. In some pressure gauge connectors, screw in the pressure gauge buffer screw or pressure gauge buffer cap at the outlet of the pressure guide hole. Since the pressure gauge buffer hole is small, if this hole is blocked by debris or there is no buffer hole When the pressure is naturally increased, the pressure gauge pointer will not move. In this case, as long as the pressure gauge buffer screw is unscrewed, the pressure can be judged again.

4, the large hole of the pressure gauge spring tube or the pressure gauge's welding joints leaks obviously, which is also the reason why the pressure gauge pointer does not move when pressurized. This kind of failure is relatively easy to judge. If the working medium is a gas, you can hear the obvious leak sound of the pressure gauge when pressurizing; if the working medium is a liquid, when you pressurize, you can see clearly from the leak point of the pressure gauge. Liquid is flowing out or ejected.

5, if the needle gauge of the pressure gauge pointer is separated from the needle piece; the needle gauge of the pressure gauge is separated from the core of the movement; the fan teeth of the pressure gauge's movement are entangled with disordered hairsprings, which may also cause the pressure gauge pointer to be pressurized Do not move.