Debugging method and self-examination skills of electric contact pressure gauge
Dec 18, 2019
Debugging method and self-examination skills of electric contact pressure gauge
Electric contact pressure gauges are widely used in petroleum, chemical, metallurgical, power station and other industrial sectors or in supporting mechanical and electrical equipment to measure the pressure of various fluid media without explosion danger. Briefly introduce the debugging and self-testing skills of the pressure gauge: Electric contact pressure gauges are usually used in conjunction with corresponding electrical devices (such as relays and contactors).
The instrument is composed of a measurement system, an indicating device, an electrical contact component device, a housing, an adjustment device, and a junction box (plug socket).
The working principle of the meter is based on the fact that the spring tube in the measurement system generates displacement under the pressure of the measured medium, and the measured value is indicated on the dial by the gear transmission mechanism. At the same time, the pointer of the meter drives the contact with the fixed contact (upper or lower limit) on the set pointer to switch on (or cut off), so that the circuit in the control system can be disconnected or switched on to the automatic control and signal Purpose of the alarm.
1. Install the assembled pressure gauge on the pressure calibrator, and connect the matching digital display ohmmeter to the junction box. Then, the pressure indication of the pressure gauge is tested for forward (boost) and reverse (boost) respectively. The error of the indication should meet the requirements of the verification regulations. While verifying the pressure indication, observe the change of the variable resistance value on the digital ohmmeter. When observing, pay attention to the resistance value output at the same time can be read (this is the same as JJG52-1999 about the requirements of reading the error value of the pressure signal of the charged contact signal device in principle). The change of resistance is the same as the change of the pressure indication. It should be linear and pay attention to its continuity. At the upper and lower limits of the pressure indication, you should pay attention to whether the variable resistor is within the effective working range, otherwise you need to adjust it by changing the position of the contacts.
2. Check whether the conductive performance between the moving contact of the variable resistor and the winding resistance is good, because it has a direct relationship with the function of the connected line. First check the contact pressure between the moving contact and the winding resistance. It should not be too small, that is, the two cannot be too loose. If the contact is too loose, it should be adjusted by changing the shape of the moving contact. Then check the contact pressure between the moving contact piece and the winding resistance, which should not be too large, that is, the two should not be too tight. If the contact is too tight, you can use cotton fiber dipped in alcohol to clean the electric contact layer of the moving contact piece. In addition, increasing the contact pressure can also reduce the contact resistance relatively.
3. The geometry and raw material performance of the variable resistor moving contact attached to the fan-shaped wheel of the remote pressure gauge should meet the requirements for easy adjustment. Otherwise, on the one hand, it affects the conductive performance between it and the winding resistance; on the other hand, it is easy to make the moving contact slide out of the effective working range of the winding resistance, thereby causing the circuit of the variable resistor to be disconnected. Installation precautions of electric contact pressure gauge
1. The electrical contact pressure gauge should be installed vertically, and strive to maintain the same horizontal position as the measurement point.
2. After the electrical circuit of the electrical contact pressure gauge is connected, the compression nut should be tightened, carefully checked and a trial operation should be performed.
3. The electric contact pressure gauge can be used up to 3/4 of the upper limit when measuring stable pressure; it can be used up to 2/3 of the upper limit when measuring alternating pressure; not subject to this limit when measuring negative pressure.
4. When opening the outlet box or adjusting the upper and lower limit setting range, the power must be cut off first.
5. Under normal use, the electric contact pressure gauge should be inspected regularly and at least once every six months. If there is a sudden problem with the pressure gauge of the electrical contact (the spring tube leaks due to leakage; the contact is welded or the reliability of the contact is turned on or off due to severe oxidation; the pointer is loose or the indicator fails), it should be immediately Repair or replace.
6. The electric contact pressure gauge should always be kept dry and clean during use.
7. Whether the explosion-proof performance of the electric contact pressure gauge is reliable or not depends mainly on the pressure strength of the flameproof enclosure and the quality of the flameproof face. Therefore, in the process of installation, use and maintenance, pay attention to the flameproof surface, and do not scratch or scratch it, especially when disassembling or disassembling the junction box of the electrical contact pressure gauge or paying special attention to it. Paint is not allowed on flameproof surfaces.
8. During maintenance, if the user is not clear about the basic knowledge of flameproof electrical equipment, and does not know the structural characteristics and requirements of flameproof electrical contact pressure gauges, do not assemble or modify it at will, otherwise it will damage the electricity. Explosion-proof performance of contact pressure gauge.